Older Posts


Tech Tip: To put the GE telemetry boxes into a 2 minute standby (e.g. to change electrodes) at MRMC, IRMC, BRMC, and IRMC (Mt. Clemens has a different procedure) press the two buttons on either side of the transmitter simultaneously for 2-3 sec until the Procedure Alarm Silence (PAS) light is flashing, then push the PAS button on the transmitter and hold until the flashing stops. Alarms are now in 2 minute standby.